Color density control uses the density sensor equipped with the RU option (Relay Unit RU-511) to measure the patch density of the chart for γ correction and perform calibration.
Instead of the γ correction by the internal IDC sensor, the density is corrected by using the printed chart image. This enables secondary transfer, fusing, and correction with each paper's characteristics taken into consideration.
Manual calibration is unnecessary. Note that in that case, on the touch panel of the main body, press [Adjustment] - [Color Density Control] - [Basic Setting], and then set [Periodical Adj. Execution] to [ON].
The color density control consists of the following three adjustments:
Details of color density control | Description |
---|---|
Basic Setting | You can enables the color density control and set the execution timing, etc.
|
Adjustment of color sensor for each paper type | You can measure and register the density adjustment data corresponding to each paper type.
|
Color density manual control | You can manually adjust the color density. |
• Image quality adjustment flow and the position of this adjustment
In the pre-output image quality adjustment flow, this adjustment is positioned shown below.
Perform the previous adjustment before proceeding with this adjustment.
To not continuously perform the series of adjustment of the image quality adjustment flow, perform Gamma Automatic Adjustment (Adjustment Flow When Changing the Screen / Regular Adjustment Flow (High Accuracy)), and then resume from the next adjustment.
• For Image Controller IC-602:
Image quality adjustment flow | Position of this adjustment | ||
Previous adjustment | This adjustment | Next adjustment | |
Adjustment flow when changing the screen | (STEP 3: Density balance adjustment) | STEP 4: This adjustment *: Optional | (STEP 5: Color balance adjustment) |
Regular adjustment flow (Standard) | This adjustment (Basic settings)
| ||
Regular adjustment flow (High accuracy) | (STEP 3: Maximum density adjustment) | STEP 4: This adjustment (Basic settings / Adjustment of color sensor for each paper type (*) / Color density manual control) *: Optional | - (None) |
• For Image Controller IC-308 or Image Controller IC-310:
Image quality adjustment flow | Position of this adjustment | ||
Previous adjustment | This adjustment | Next adjustment | |
Adjustment flow when changing the screen | (STEP 3: Density balance adjustment) | STEP 4: This adjustment *: Optional | (STEP 5: Controller calibration) |
Regular adjustment flow (Standard) | This adjustment (Basic settings)
| ||
Regular adjustment flow (High accuracy) | (STEP 3: Maximum density adjustment) | STEP 4: This adjustment (Basic settings / Adjustment of color sensor for each paper type (*) / Color density manual control) *: Optional | - (None) |
You can enables the color density control and set the execution timing, etc.
When using the color density control function (Adjustment of color sensor for each paper type / Color density manual control), always set the [Use Color Density Control] to [ON].
On the touch panel of the main body, press [Adjustment] on the [MACHINE] screen.
The [Adjustment Menu] is displayed.
Press [Color Density Control] on the [Adjustment Menu].
Press [Basic Setting].
Make a setting for each item.
Make settings according to your environment. The settings in this case are as follows:
Item | Description |
---|---|
[Use Color Density Control] | [ON] |
[Periodical Adj. Execution] | [ON] |
[Adjustment Execution Timing] | [Before Job Start] |
[Chart Tray Setting] | [OFF] |
[Switch Color (1st or 2nd)] | [Precise] |
To automatically perform the density control with the RU option (Relay Unit RU-511), set the [Periodical Adj. Execution] to [ON].
You can measure and register the density adjustment data corresponding to each paper type.
* Perform optionally this adjustment as needed.
To measure the chart, use the X-rite Measure tool (for the spectrophotometer i1iSis or i1Pro) or Color Care measure (for the spectrophotometer FD-5 BT). Install the tool in advance.
You can download the X-rite Measure tool free of charge at the following Web site. After installing the tool, restart the computer.
ProfileMaker 5 v5.0.10:
http://www.xrite.com
To use the spectrophotometer, i1iSis, perform Step 13 (select [Measure] from the menu) followed by the procedure of "• For the spectrophotometer, i1iSis."
To use the spectrophotometer FD-5 BT, perform Step 8 (output the color chart) followed by the procedure of "• For the spectrophotometer FD-5 BT."
On the touch panel of the main body, press [Adjustment] on the [MACHINE] screen.
The [Adjustment Menu] is displayed.
Press [Color Density Control] on the [Adjustment Menu].
Press the [Adj. of Col. for Paper Type] of [Expert Setting].
Select the number where the output density control value is to be registered and press [Print Mode].
If you select the registered number, the adjusted value is overwritten.
Enter the registered name of the color control for each paper type, and press [OK].
On the [PRINT MODE] screen, press [Quality Adj.].
Press of [Screen Pattern]. After the selection, press [Close].
After selecting the tray on the PRINT MODE screen, press Start on the control panel.
To set the output tray, press [Output Setting] at the upper right of the PRINT MODE screen.
The color chart is output for measurement using the spectrophotometer.
Start the computer where the X-rite Measure tool has been installed, select [Start] - [All Programs] - [X-Rite] - [ProfileMaker Pro 5.0.10] - [MeasureTool].
X-rite Measure tool starts.
The following description uses screen examples in English. To change the display language for the menu and messages, select the language from [Language] on the [Measure Tool] toolbar, and then restart the X-rite Measure tool.
When using the spectrophotometer FD-5 BT, perform the measurement on the computer installed with Color Care measure. For the subsequent procedure, refer to "• For the spectrophotometer FD-5 BT."
When the following message is displayed, click [OK].
Connect the instrument to the computer. From the menu, click [Configuring].
The [Instrument Configuration] screen is displayed.
In [Instrument], select the spectrophotometer to use. When [OK] is displayed, close the screen.
Example: This example selects [Eye-One Pro] (i1Pro).
From the menu, click [Measuring].
The subsequent operations are described for the case where the spectrophotometer i1Pro is used. For measurement using the spectrophotometer i1iSis, refer to "• For the spectrophotometer i1Sis."
From [Test Chart], select [Custom].
Specify the [Columns] and [Rows], and click [OK].
Specify them according to the size of the chart.
For the spectrophotometer i1Pro / Large size (3 sheets of checking charts): [Columns] = 4, [Rows] = 11
For the spectrophotometer i1Pro / Middle size (4 sheets of checking charts): [Columns] = 4, [Rows] = 8
Click [Start].
When the following message is displayed, horizontally place the spectrophotometer on the calibration dock and click [OK].
Stack 10 sheets of blank paper whose type is the same as that of the paper where chart has been printed, and place the printed chart on top of it.
To eliminate the influence of the color of the table, place 10 sheets of paper of the same type as the chart underneath the chart to be measured.
When the following screen is displayed, place the spectrophotometer on the patch at the top in the upper left, and press the button for the spectrophotometer. When beep sounds, check that the read patch is displayed on the screen.
Continuously, place the spectrophotometer on the next right patch, and press the button for the spectrophotometer. Measure all the patches in the same operation.
Measure the rest of the patch rows by sliding the instrument the same way.
If you place the spectrophotometer on the wrong place (patch), click the button to go back to the previous patch position.
When you are finished with the chart measurement, click [Close].
Each time the chart measurement is finished, the message is displayed to prompt you to feed the next chart. Save the measurements for each sheet.
Continuously, go to "Saving the measured data," and save the measurement data.
• For the spectrophotometer i1iSis
To use the spectrophotometer, i1iSis, perform Step 13 (from the menu, select [Measure]) followed by the procedure below.
From [Test Chart], select [Open].
Select the script data for measurement according to the paper size, and click [Open].
For details about the script data, contact your service representative.
Specify them according to the size of the chart.
• OutdensenA3V01_S1Cut.txt: Large-size chart (3 sheets of checking charts)
• OutdensenA4V01_S1Cut.txt: Middle-size chart (4 sheets of checking charts)
• OutdensenB5V01_S1Cut.txt: Small-size chart (5 sheets of checking charts)
On the following screen, click [Start].
When the following message is displayed, insert the chart into i1iSis.
Cut off the left side of the chart using scissors or such other tools so that the length between the left end of the chart and the black diamond symbol (♦) becomes about 10 mm.
Align the left end of the paper to the left end of the insert opening of i1iSis, and automatically feed the paper one by one. When correctly inserted, the chart is pulled in a few centimeters and comes back a little, and then after a little while, reading starts.
Let the machine automatically feed the paper one by one from the black diamond symbol.
Measurement automatically starts.
• For the spectrophotometer FD-5 BT
For the spectrophotometer FD-5 BT, use Color Care measure.
To use Color Care measure, you need to acquire a license key. For details, refer to the user's guide that comes with Color Care measure.
On the Color Care measure screen, select a chart based on the number of color chips. You should select the one having the same number of rows and columns of color chips printed on it.
When you use Color Care measure to register data through the Web, a password entry is requested. Enter the administrator password registered on the machine.
You are also asked to enter 4-digit number printed on the chart (2 digits of Each Paper Type Color Adj. No. + 1-digit total page count + 1-digit page number) and the IP address for the machine.
You can save the data of the output density measured with the spectrophotometer to a USB device such as a USB memory stick or to the computer on the network used by the machine.
From the [File] menu, select [Save As].
If you use [Export Lab], an error occurs when registering the color adjustment data. When saving the data, always use [Save As].
Specify the destination to save.
Specify USB memory or the computer connected to the machine via the network.
Destination to save | Description |
---|---|
USB Memory | Connect the USB memory to the computer. • C1100: Create a folder, "\C1100\ADJUST_DATA," in the root (immediately below) of the USB memory, and specify the place as the destination to save. |
Computer on the network | Open the computer connected to the network used by the machine, and specify a desired folder. |
Enter the file name to save the measured data and click [Save].
Specify the "4-character identification number printed on the chart."
Example: 0431.txt, 0432.txt, 0433.txt
The measured data of the color density control is saved.
Repeat the same process to save all the measured data.
You can register the measured data (color density value for each paper type) saved in the USB memory or the computer on the network into the machine.
Place to save measured data | Operation |
---|---|
Measured data saved in the USB memory | Connect the USB memory and register the data using the touch panel of the main body. Refer to "• Registering the measured data saved in the USB memory." |
Measured data saved in the computer on the network | Send (upload) the measured data using Web Utilities from the Web browser and then register the data. For details about Web Utilities, refer to Overview of Web Utilities. Refer to "• Registering the measured data saved in the computer on the network." |
• Registering the measured data saved in the USB memory
On the touch panel of the main body, press [Adjustment] on the [MACHINE] screen.
The [Adjustment Menu] is displayed.
Press [Color Density Control] on the [Adjustment Menu].
Press the [Adj. of Col. for Paper Type] of [Expert Setting].
Select the number that was registered when outputting the chart, and press [USB Memory].
For the registered number, [Waiting measured data loading] is displayed.
The following screen is displayed. Connect the USB memory where the measured data has been saved to the main body, and press [OK].
Select the file name of the measured data to be registered and press [OK].
Multiple files corresponding the checking chart are selected at a time (you cannot select the files individually).
The adjusted value of the output density is registered in the machine.
• Registering the measured data saved in the computer on the network
On the computer on the network where the measured data has been saved, open a Web browser.
It is recommended that you log in to Web Utilities from the computer where the measured data has been saved.
In the URL field, enter "http://IP address of the machine (or the host name):30091" and press the Enter key.
If the controller of EFI is implemented, ":30091" at the end of the IP address is unnecessary.
The main page of the Web Utilities is displayed.
Click on [Machine Manager Setting].
Enter the [User Name] and [Password] and click [OK].
Click on [Color Measuring data Input].
Specify the [File Name] and [File Path] and click [Upload].
File name (Example: 2021115011C4CC_i1pro_2012-10-15.txt)
Click [Reference] of the [File Path] and select the measured data saved in the computer.
Repeat the same process to upload all the measured data.
On the touch panel of the machine, press [Adjustment] on the [MACHINE] screen, and then press [Color Density Control] on the [Adjustment Menu] screen.
Refer to Steps 1 and 2 of "• Registering the measured data saved in the USB memory."
Press the [Adj. of Col. for Paper Type] of [Expert Setting].
Select the number that was registered when outputting the chart, and press [Engine Archived Data].
Select the file name of the measured data to be registered and press [OK].
Multiple files corresponding the checking chart are selected at a time (you cannot select the files individually).
The adjusted value of the output density is registered in the machine.
You can apply the registered adjusted color density to the machine.
On the touch panel of the main body, press [Paper Setting].
The Paper Setting screen will be displayed.
Select the tray to which the adjusted color density is applied, and press [Change Set].
Press [Color Density].
Press [Registered Adj. Data].
Select the adjusted value to be applied from the list and press [OK].
You can manually adjust the color density.
On the touch panel of the main body, press [Adjustment] on the [MACHINE] screen.
The [Adjustment Menu] is displayed.
Press [Color Density Control] on the [Adjustment Menu].
Press the [Color Density Manual Control] of [Expert Setting].
Press [Print Mode].
On the PRINT MODE screen, press [Quality Adj.].
Press of [Screen Pattern]. After the selection, press [Close].
After selecting the tray on the PRINT MODE screen, press Start on the control panel.
To set the output tray, press [Output Setting] at the upper right of the PRINT MODE screen.
When the chart is output, the manual control is completed.